![]() Enclosure source current, ground open Do not continuously activate for longer than one minute. Use of duty cycles greater than 25% (10 seconds active followed by 30 seconds inactive) will increase the risk that heat build-up under a return electrode may be high enough to injure the patient.Under maximum power settings and rated load conditions, the ForceTriad energy platform is capable of operating a duty cycle of 25% defined as 10 seconds active and 30 seconds inactive, in any mode for a period of 4 hours.Keying delay: User selectable in 500 msec increments from 0 sec to 2.5 sec.± 5% of Full-Scale activation impedance while keying inactive.+ 50%, -14% of full-scale activation impedance while keying active at 10 watts or below.Measurement accuracy: ± 14% of Full Scale activation impedance while keying active above 10 watts.Deactivation impedance: User selectable: 1,500 Ω 1,800 Ω or 2,200 Ω.Interrogation frequency: 80 kHz ± 10 kHz. ![]()